Back to results
Water SA Manuscript

Measurement of pH, alkalinity and acidity in ultra-soft waters

Published: 2001
Report No: 1 428,00
Authors: Loewenthal RE, Morgan BE, Lahav O

Abstract

Conventional characterisation of low alkalinity waters via pH measurement and titration of total alkalinity to a prescribed end-point invariably leads to large errors. These errors result from instability of the pH probe and an unknown titration end-point. In this paper two indirect methods (termed the “double Gran function” and the “blend” method) for the characterisation of such waters are evaluated critically. A blend composed of the raw water, sodium chloride (to increase conductivity), and standard bicarbonate (to increase buffering capacity) was titrated with standard strong acid in two pH regions: 6.3 < pH < 7.0, and 3.5 > pH < 4.0. In both methods, total alkalinity was determined using the latter set of points, and the first Gran function. In the double Gran function method the upper set of titration points was used to determine CO 2 acidity using the second Gran Function. In the “blend” method, equilibrium chemistry data were used to calculate total acidity for each point based on the known total alkalinity, pH reading, temperature and ionic strength.